The US XPD-24-200 sampler suspended in clear water in the tow tank

The US XPD-24-200 in the tow tank at the former USGS Hydrologic Instrumentation Facility, John C. Stennis Space Center, Mississippi (USGS photograph).

River sampling for suspended-sediment and water-quality work has relied on separate samplers: one that grabs water at a single depth and one that collects as it moves down and up. A USGS report describes a single sampler, the US XPD-24-200, built to do both with a collapsible bag. The Federal Interagency Sedimentation Project (FISP), which has led the development and standardization of sediment-sampling equipment since 1939, designed and tested it with Carnet Technology. A first prototype appeared in 2021 and a second, the XPD-24-200, in 2024.

What was wrong with older samplers

FISP samplers are calibrated to be isokinetic, meaning water enters the nozzle at the same speed as the stream flowing past. The measure is intake efficiency (IE), the ratio of nozzle velocity to ambient velocity, where 1.0 is perfect and FISP samplers are calibrated to 0.9-1.1.

  • The US P-6 point sampler holds at most about 475 mL, which is often too little at low sediment concentrations. It also needs a compressed-air mechanism for its rigid bottles, which adds mechanical complexity, and it takes only a 3/16-inch nozzle.
  • Depth-integrating samplers such as the US D-96 have no valve, so they fill on the way down and on the way up. That can overfill them and limits how deep they can go.
  • Doing both kinds of sampling has meant owning two samplers, which costs more and takes longer.

How the new sampler works

The sampler hangs from a cable and has a streamlined body with a tail of stabilizing fins. Inside, a 3-liter flexible bag fills under the surrounding water pressure, so no compression chamber is needed. An electric valve, controlled by current sent down the cable, opens and closes the path from the nozzle to the bag. It runs on 18 volts of direct current: 12 volts did not open the valve reliably, especially with a full reel of cable. Nozzles of 3/16, 1/4 and 5/16 inch fit.

FeatureDetail
Length36.6 inches
Weightabout 200 pounds
Construction3D-printed plastic around a core of resin-infused lead; nozzles, flow path and valve of PTFE
Unsampled zone6 inches

A key feature is unidirectional depth-integrating sampling: the valve opens at the surface, stays open on the way down, closes near the bottom and stays shut on the way up. That avoids overfilling and lets operators sample deeper or in faster water. It also needs cheaper equipment, because a reel clutch can control the descent and a simpler motor can lift the closed sampler.

The tests

The sampler hanging from a crane over Lake Tuscaloosa, Alabama

The sampler hanging from a crane on Lake Tuscaloosa, Alabama (USGS photograph).

Tow-tank testing took place over 10 days in 2023 and 2024, towing the sampler at 1.5 to 12 feet per second in water up to 10 feet deep. Lake testing took place over 3 days in April 2025 on Lake Tuscaloosa, with the sampler towed from a motorboat at 1.43 to 15 feet per second in water more than 90 feet deep. Tests covered point and depth sampling, volume, leaks, overfilling and valve operation. Results from 2023 are not reported because early modifications strongly shaped them.

What they found

  • The largest sample was 2,624 mL, 5.5 times the P-6's maximum.
  • About 73 percent of IE measurements fell within 0.9-1.1. Around 24 percent were below 0.9, mostly at velocities under 4 feet per second, and about 2 percent were above 1.1.
  • In the tow tank, mean IE ranged from 0.926 to 0.952 across nozzles. In the lake, it ranged from 0.834 to 0.946 and was more scattered, which the report links to natural variability such as turbulence and temperature gradients.
  • The sampler showed stable performance and reliable valve operation and was simpler to deploy and maintain than older samplers.

Further tests are needed, including with a perfluoroalkoxy (PFA) bag, to show the sampler meets the water-quality sampling protocols in the USGS National Field Manual. The report also suggests long-duration valve cycling, flume tests and field comparisons of sediment concentration with approved FISP samplers.

Sources

コメント

Spinner Logo

自然 のほかのページ

Hub Nexus editors 作閲覧数: 150
Are you a traveller to plan your next destination? Or have you not had any travels at all for quite some time with overwhelming daily schedules? It might be a good time to take a break to air it out. Here're some of the iconic places in the world to consider for your next destination.
Hub Nexus Nomad 作
From the glowing waves of the ocean to fireflies lighting up the night sky, bioluminescence is one of nature’s most mesmerizing phenomena. This article dives into the science behind bioluminescence, its role in nature, and how this glowing phenomenon has inspired both innovation and awe in the human world.
CC0 1.0(パブリックドメイン)
FEQinput helps hydrologists read, search, edit and run input files for FEQ, the USGS model of unsteady flow in open channels, showing descriptions of input lines to help new users overcome a steep learning curve.
CC0 1.0(パブリックドメイン)
The United States Congress designated the Cottonwood Canyon Wilderness in 2009 and it now has a total of 11,667 acres. All of this wilderness is located in Utah and is managed by the Bureau of Land Management. The Cottonwood Canyon Wilderness is bordered by the Cottonwood Forest Wilderness to the no
CC0 1.0(パブリックドメイン)
Sea ice grows throughout the winter and melts throughout the summer, reaching its maximum extent in late February or March, and its minimum extent in September. The ice that survives at least one summer melt season is typically thicker and more likely to survive future summers. Since the 1980s, the
CC0 1.0(パブリックドメイン)
Alaska's Deep-Sea Sponges: Building a Library of Species Nobody Has Named Yet