Images from different MRI machines, or from the same machine months apart, have been hard to compare, which matters in clinical trials that run at several sites. The answer is a standard reference object, called a phantom: a stand-in for a body section or tissue type that is imaged at regular intervals so that results can be compared across machines and across time.
NIST's Biomagnetic Imaging Standards project, led by Stephen Russek in the Physical Measurement Laboratory (PML), has developed standard phantoms over the past few years. As a result, "quantitative MRI", with rigorous metrics and identified sources of error, is becoming recognized as vitally important. Russek says most of the basic measurement issues are resolved, among them tumor volume over time, resolution, spatial distortion and relaxation times. The team recently acquired a 7 tesla small-bore scanner, big enough to characterize every component of the phantoms it makes for clinical use. Ordinary clinical MRI systems run at 1.5 T or 3 T.
The need for standards is growing, Russek says, as researchers devise new procedures and imaging goals.
Diffusion, flow and temperature
- Diffusion. One new phantom is optimized for imaging diffusion, the way water moves in the body. Diffusion is inherently affected by temperature, and temperature varies from scanner to scanner, so the team needed an MRI "thermometer".
- Flow. Ron Goldfarb, who leads the Magnetics Group in PML's Electromagnetics Division, says some MRI modes are not fully used partly because they have no standards. They can track changes in adjacent voxels (the three-dimensional counterpart of pixels) over time to show fluid motion. One example is water flowing along nerve tracks in the corpus callosum, which connects the two cerebral hemispheres. The next big thing, he says, is a phantom that measures perfusion, diffusion and flow.
- Temperature. Russek says temperature matters for perfusion and diffusion studies and for procedures such as RF ablation and focused ultrasound heating, where MRI guides the focus and images the local temperature. The damage threshold for short exposure is near 50 °C, and temperature can vary widely from place to place. Quantitative data on temperature inside the body does not really exist yet. Project member Michael Boss is developing a local thermometer that uses a rare-earth chelate as a reference. Its paramagnetic spin is sensed by nearby protons and depends strongly on temperature.
Other new work includes a prototype standard breast phantom, ways for MRI makers to test the accuracy of their software, and low-field MRI, which uses fields as weak as the Earth's (about 50 microtesla). Russek says it can be cheaper because it needs no large superconducting magnet, and it may give better contrast for some tissues, such as prostate cancer. It has not been fully validated, and part of the team's work is to help validate those measurements.
A breast phantom for a clinical trial

Design for the prototype breast phantom (NIST).
Katy Keenan says an estimated $16 billion a year is spent on breast cancer treatment in the United States, and several clinical trials are testing treatment protocols, with quantitative MRI increasingly used to measure outcomes. In October 2012 NIST partnered with Dr. Nola Hylton to develop a breast phantom for the ISPY-2 trial, together with the American College of Radiology Imaging Network (ACRIN) and the National Cancer Institute's Quantitative Imaging Network.
ISPY-2 tests new breast cancer therapies at more than 20 sites nationwide. Its goal is to validate MRI measurements, rather than invasive biopsy, as the way to judge response to treatment, so the data must be comparable across sites.
Breast MRI is hard because of the uneven mix of fat and fibroglandular tissue, which causes image artifacts, and because breasts differ in size. Keenan lists three goals for the phantom:
- Include fat and fibroglandular components suited to the ACRIN trial's MRI measurements: tumor volume, proton-spin T1 relaxation time and diffusion.
- Include SI-traceable components like those in the NIST system phantom.
- Work with every breast MRI coil on the market.
At the time of writing, the team was finalizing the prototype design and expected to finish initial imaging of it by the end of the summer. Industry and other trial groups had already shown interest, and production needs to be scaled up.
How MRI makes an image
In an MRI scanner the strong magnetic field lines up many of the protons in the hydrogen atoms of the body's water. A radio-frequency pulse "flips" their orientation and is then switched off. The scanner detects the times (T1 and T2) the protons take to realign with the field. These relaxation times differ by tissue type, among other factors, and software converts the differences into a grayscale image.