Until now, the most reliable way to measure pressure has used a technique invented almost 400 years ago: a column of mercury. NIST researchers now report that a mercury-free sensor that measures pressure with light beats the U.S. mercury standard at low pressure.
The old way and the new
A mercury manometer reads pressure from the height of a mercury column. It is slow, about a minute per measurement, and bulky: the U.S. primary pressure standard is a 3-meter (about 10-foot) manometer that reaches through the ceiling of its lab. Mercury is a toxin whose use is being phased out of products and manufacturing internationally.
The Fixed Length Optical Cavity (FLOC) is a chamber about 15.5 cm long and 5 cm square, a fraction of that size. It compares the optical path length between pairs of mirrors on two channels, one in vacuum and one filled with ultra-high-purity nitrogen. As gas pressure rises in the nitrogen channel, light takes longer to bounce between the mirrors, so the channel behaves as if it were longer. Laser frequency must change to keep a standing wave, and comparing it with the vacuum channel gives the gas's index of refraction, from which pressure follows. (Pressure does slightly distort the channels, but the team measured and characterized that noise.) Early results had shown the FLOC to be faster and higher in resolution than mercury. Jay Hendricks, PML's Thermodynamic Metrology Group Leader, calls it the biggest fundamental change in pressure measurement since the mercury manometer was invented in 1643, and says the technique can cover up to nine decades of pressure, which now takes six primary standards.

Closer look at the FLOC cavity.
The head-to-head test
Jacob Ricker connected a FLOC and the manometer to the same pressure source. The two could not share a gas sample, because the FLOC needs ultra-pure nitrogen, so a capacitance diaphragm gauge, which measures the difference between two pressures, linked them. Each FLOC measurement takes under a second. The comparison showed the FLOC more precise than the manometer at pressures below about 1 kilopascal (kPa).
To check this, the team built a second FLOC and tested all three against each other:
| Finding | Result |
|---|---|
| Agreement between the two FLOCs | within about 1 part per million |
| Disagreement of manometer with either FLOC | up to twelve times worse |
| FLOC reproducibility and linearity, 100 Pa to 100,000 Pa | about 1 part per million |
| FLOC accuracy above 1 kPa | two times less accurate than the manometer |
Because both FLOCs were linear, the team blamed the disagreement on nonlinearities in the manometer.
Patrick Egan credits part of the agreement to baking the FLOCs for two months at 373 K (100 degrees C) in a vacuum, which cures the bonding that attaches the mirrors so they stay stable. Afterward the mirror pair on each channel drifted only about 6 femtometers per hour, which Egan said is smaller than the nucleus of a gold atom.
Why the FLOC lags at higher pressure, and what is next
The FLOC's pressure reading depends on how accurately the refractive index of nitrogen is known, and that is limited by how well pressure can be measured with the best instrument available, the mercury manometer. To fix that, the team is building the variable-length optical cavity (VLOC), a cousin of the FLOC using helium, to measure nitrogen's refractive index better than has been done. In the near term, the FLOCs will be used with the manometer to calibrate customers' pressure gauges.
Sources
- National Institute of Standards and Technology, "Head to Head: Photonic Pressure Sensors vs. Mercury-Based Standard" by Jennifer Lauren Lee, https://www.nist.gov/news-events/news/2016/04/head-head-photonic-pressure-sensors-vs-mercury-based-standard; rewritten in hubnx's own words.