The tool that semiconductor scientists trust to measure chip features only tens of nanometers wide has a problem: its own tip has to be measured first, and the tool is one of the only ways to measure something that small. A team at the National Institute of Standards and Technology (NIST) found a way around it with a reference sample, and an improved version now allows tip calibration with an uncertainty under 1 nm.

Scanning electron microscope image of the SCCDRM sample, with near-vertical sidewalls of varying linewidth

Figure 1. Scanning electron microscope image of the SCCDRM sample, showing near-vertical sidewalls with varying linewidths, high uniformity and known linewidth values.

Why the tip matters

As chip features shrink, visible-light microscopes can no longer examine them. The critical dimension atomic force microscope (CD-AFM) measures them by dragging a tip of roughly 10 nm across the surface, like a record player needle in a groove. Its potential width uncertainty is as low as 1 nm, which makes it the tool of choice in the semiconductor industry.

But raw CD-AFM data show linewidths that include the width of the tip, which is then subtracted. A user who believes the tip is 10 nm wide when it is actually 15 nm gets measurements with unacceptably large uncertainty. That is why NIST's Ron Dixson says linewidth calibration and tip-width calibration are, with an AFM, basically the same thing.

The 2005 standard

Dixson and colleagues worked on the problem periodically for 10 to 15 years. The first NIST standard for tip calibration came in 2005, with a tip width uncertainty of approximately 1 nm. It grew out of a collaboration with SEMATECH that sent Dixson there on assignment for three years from late 2001: NIST had no CD-AFM and no AFM linewidth capability, and instruments like that cost over a million dollars at the time, while SEMATECH had one and no one to run it.

The team used the CD-AFM and a transmission electron microscope (TEM) together. TEM is destructive, but its images let researchers count the atomic planes of a sample (strictly, interference fringes that correspond one to one with the planes). The nondestructive CD-AFM served as a comparator between the samples sacrificed for TEM and those sent to users, tying the results to the SI meter. The finished product was a 1 cm² piece of silicon with multiple measurement targets in the center of a 200 mm wafer, distributed to SEMATECH member companies.

Before that, researchers used commercial products, some of which tended to damage tips, and those offered an uncertainty of about 5 nm at best. Others used production samples with known values, with about the same 5 nm. In practice only a few of the 2005 chips had the full 1 nm uncertainty; most were closer to 1.5 or 2 nm because their linewidths were not quite uniform along the features. The NIST master sample was known to 0.6 nm (k = 2), but transferring it to the distributed samples added uncertainty.

What changed

Scanning electron microscope image of a CDR50 type CD-AFM tip

Figure 2. Scanning electron microscope image of a CDR50 type CD-AFM tip. These tips have a nominal geometric width of 50 nm, but the width must be calibrated for each tip.

The new reference sample is a single crystal critical dimension reference material (SCCDRM), as the 2005 sample was, made with a lattice-plane-selective etch on (110) silicon. A pattern of near-vertical sidewalls of varying linewidth, high uniformity and known values is etched onto the chip. A user measures the sample, compares the result to the known value and adjusts the tip width entered in the instrument as necessary.

The improvement, which Dixson puts at up to a factor of five when the non-uniformity of the older samples is taken into account, came from swapping the etchant: potassium hydroxide (KOH) instead of tetramethyl ammonium hydroxide. Part of fabrication is removing an oxide hard mask from the chip, and KOH removed as much of it as possible without damaging the narrow sidewalls. On the 2005 samples the sidewalls were not uniform along their length, so a tip not placed at a precisely defined spot could give a wrong reading. The new samples are more consistent chip to chip and offer an uncertainty under 1 nm.

Cross-check and next steps

Dixson also took part in a comparison with PTB, Germany's national metrology institute, which was developing its own calibration solution for CD-AFMs. The two institutes cross-checked their samples, and the not-yet-published results showed very similar uncertainty. The new samples had not been widely distributed, and a wider distribution was under consideration. NIST was also discussing further calibrations, such as the vertical edge height, the distance from the base of the tip to the point that touches the sample, for which no official standard exists.

Sources

Комментарии

Spinner Logo

Ещё на тему «Наука»

Автор: Hub Nexus editors
Top 10 in AI, week of 12 October 2026
Автор: Hub Nexus editors
Top 10 in AI, week of 2 November 2026
Автор: Hub Nexus editors
Top 10 in AI, week of 26 October 2026
Автор: Hub Nexus editors
Top 10 in AI, week of 19 October 2026
CC0 1.0 (общественное достояние)
Laser therapy uses an intense, narrow beam of light to remove or destroy cancer and abnormal cells that can turn into cancer. Learn how lasers work and about the conditions that they are used to treat.
CC0 1.0 (общественное достояние)
From a Networking Event to a Seed Study: Tracking Eelgrass Flowers in Barnegat Bay