Absorbance spectroscopy identifies even small amounts of a substance, such as DNA or an antibody, by how it absorbs light. A biotechnologist developing new medicines or a better test for forensic analysis might use it frequently.
The reference material
NIST's Material Measurement Laboratory (MML) recently developed a new Standard Reference Material, SRM 2082, for samples where only a little material is available. It consists of two solutions: tryptophan, an amino acid, and uracil, a nucleic acid found in RNA. Its certified property is an equation relating the pathlength of light through the solutions to their absorbance.
Before release, MML needed an ultra-accurate figure for the width of the NIST cuvettes, the tiny containers used in certification, because that width sets the pathlength light travels through the solution. It asked dimensional metrologists in NIST's Physical Measurement Laboratory (PML).

A cuvette with a 2 mm gap length, lit by a laser spot from the triangulation probe.
Why the usual tool failed
Each miniature cuvette is nominally 45 mm long by 10 mm wide, smaller than a typical thumb drive, with three possible pathlengths: 2 mm, 1 mm and 0.5 mm. Geometric imperfections make the gap between the walls vary slightly from place to place, so each cuvette's gap needed careful characterization, says PML's Wei Ren.
The gold standard is a high-accuracy Coordinate Measuring Machine (CMM), whose hard probe tip touches the surface. Here the problem was the cuvette's large length-to-width ratio. Ren had tips of the right size, but smaller tips need shorter shafts to stay stiff, and those shafts were too stubby to reach the bottom of the containers.
The workaround
Ren often used laser triangulation, a non-contact method, on ceramic and other opaque surfaces. A laser spot lands on the object, a lens collects the reflected light and focuses it onto a CCD, and the spot's position on the CCD changes with distance to the object. She had never used it on transparent glass. Her fix was to aim the laser at a 20 degree angle rather than straight down. Light reflected back to the detector from each of the four air-glass interfaces, and a program computed the distances between them, including the gap between the two inner walls.
She probed each cuvette's gap at 6-mm intervals. Because these were not absolute measurements, she checked at least part of each cuvette against a standard touch probe on the NIST M48 CMM.
Her measurements gave uncertainties under 1 micrometer. That let the MML team state an expanded uncertainty for the materials' absorbance, as a function of pathlength, of less than 0.08 percent.
Eric Stanfield, of PML's Dimensional Metrology Group, said the team spent a lot of time failing to find a complete probing solution. Ren's approach, combining probing with a non-contact technique, "worked remarkably well." The new SRM was expected to be released by April 2016.