Measuring the color of a car's finish, how much light a tinted window lets through and how an Earth-observing satellite's sensors are calibrated all depend on accurate measurements of reflected and transmitted light. The National Institute of Standards and Technology (NIST) sets and maintains those scales for the United States, and it completed major upgrades to the two key instruments at their core.
Operators of reflectance and transmittance instruments calibrate their equipment against the scales, much as tuning forks provide a reference for piano notes. That ensures different devices produce comparable data. NIST maintains the scales in the near ultraviolet (wavelengths above about 250 nanometers), the visible (400 to 700 nm) and the near- and short-wave infrared. Disseminating those scales, improving measurements traceable to the SI, supporting U.S. defense labs and coordinating with national metrology institutes worldwide are major goals of the program.
The measurements matter well beyond color. Reflected light shapes how we see clothing, automobile finish, paint, coatings and food, and accurate measurement of reflected and transmitted light, including wavelengths we cannot see, is also essential in manufacturing, pharmaceuticals, medical imaging, printed materials, optics and photonics, and satellite observation of weather, crop health and ocean conditions.
Reflectance: ROSI
How reflected light looks depends on the angle at which light strikes a sample, the angle at which it is detected, the spectrum of the light source and the nature of the surface. NIST has long been a world leader in measuring those effects, and one important example is the newly upgraded Robotic Optical Scattering Instrument (ROSI) in its Physical Measurement Laboratory, described in the October 2021 issue of Applied Optics.

Credit: Sean Kelley/NIST.
ROSI can measure reflection from specular (mirror-like) samples and diffuse ones, rough surfaces that spread out the reflected light. Its illumination and viewing angles each range from 0 to 80 degrees. An exceptional feature is that it can measure at out-of-plane viewing angles, rotated away from the incident plane, as when the sun rises in the east over a lake and an observer views the water from the north shore. Other national metrology institutes have out-of-plane reflectance instruments, but NIST says it is unique in providing such measurements over so wide a range of wavelengths. ROSI's customers include commercial interests and national and international government laboratories, according to NIST scientist Heather Patrick.

Credit: Sean Kelley/NIST.
- Satellites. Earth-monitoring satellites use different wavelength bands of reflected light to determine ocean temperature, chlorophyll concentration, sea-surface height, the speed and direction of currents and more. Their instruments need a reference to calibrate against, which can be a diffuser with exactly known reflectance and measurements of the sun's angle on it. NASA's Diffuser Calibration Laboratory is one of ROSI's longest-running customers, and NASA relies on ROSI calibrations for the reflectance values assigned to its solar diffusers. Patrick notes that a conventional weather satellite has been estimated to cost between $3 billion and $5 billion to develop and launch, so accurate measurement of its components is critical.
- Other uses. Reflectometry can distinguish oxygenated from de-oxygenated blood, healthy from non-healthy plants, and identify the concentration and chemical makeup of aerosols in the atmosphere. Makers of color-measuring instruments rely on standards measured on ROSI; a white sample, whose reflectance varies little from one wavelength to the next, gives a known top of the scale for calibrating spectral response.
- A longer reach. ROSI has been in continuous development since 2013 and has taken over many measurements from NIST's aging STARR instrument, which is still used for hemispherical reflectance. A shorter-wavelength light source was added: the original could provide tunable light down to about 430 nm, visible deep violet, and the new one extends that to 250 nm in the UV. One reason is the need to measure the reflectance of the UV sources used in germicidal applications. Patrick says UV light can sterilize air and surfaces, for example in health care settings, and UV reflectance data can help engineers design more efficient lamp housings and help determine the UV dose needed to inactivate pathogens.
Transmittance: a faster, sharper UV service
NIST's ultraviolet transmittance calibration service, which assesses how much light of different wavelengths passes through an object, gained a lab upgrade that achieved a 70% reduction in uncertainty and halved the measurement times. The service is for customers who work with materials that filter light: glass manufacturers optimizing window coatings, builders of remote sensing instruments that rely on filters, pharmaceutical companies and military standards laboratories. It also supports consumer uses indirectly: police might use a commercial transmittance sensor, traceable to NIST or another national metrology institute, to check how much light passes through a tinted car window and whether it complies with local laws.
The service covers wavelengths from the UV at 250 nm up into the shortwave infrared, but the improvements matter most between 250 and 450 nm. The setup can measure neutral density filters, samples of tinted glass designed to absorb light of various wavelengths, which customers use to assess the performance of their own transmittance devices. To calibrate a sample:
- A broadband source, one that produces many wavelengths, sends light through a monochromator, a series of gratings and mirrors that selects a single wavelength.
- That light passes through the customer's sample and reaches a detector.
- For each wavelength the researchers also remove the sample and send the beam straight to the detector, which gives the light's intensity before and after transmission.
Two changes made the difference. The old UV source was a deuterium lamp, whose fairly weak beam limited the signal. It was replaced by a laser-driven light source, in which a high-power laser makes a gas-filled bulb emit broadband light. That source gave more light but was unstable, with the UV output fluctuating over a couple of minutes, so the team added a monitor line that measures the light coming out of the monochromator in real time. A change in protocol let the team make the calibrations twice as fast: instead of several scans per spectrum, it takes more readings within a single scan, according to Catherine Cooksey.
With the speedup, the main reference instrument, which makes measurements traceable to the SI, can be used for customer work directly; before, the time it needed was so long that a commercial instrument was used instead. As of the story, the new system was about to be used for customer measurements, and the researchers were preparing a formal comparison with a similar setup at the National Research Council Canada, which would let both labs validate their methods and tie them to world measurements.
Sources
- National Institute of Standards and Technology, Measuring Up: Light reflection and transmission, reported and written by Curt Suplee and Jennifer Lauren Lee; rewritten in hubnx's own words. The source gives two different upper limits for the infrared (about 2400 nm for the scales and 2500 nm for the transmittance service), so neither is stated.