Principle of volume displacement measurement applied in the calibration of a meter. Figure from NIST's article.
Over the past few years the United States has imported, and exported, about 8.5 million barrels of petroleum a day. At $75 a barrel, the price of crude oil on March 1, 2023, that is roughly 230 billion dollars of petroleum traded each year, so a 1 percent error in measuring it would be worth 2.3 billion dollars a year, the National Institute of Standards and Technology (NIST) points out. Accuracy matters. But a meter in a production pipeline has to be checked regularly without removing it, stopping the flow of product or halting the metering. The tool for that job is a volume displacement prover.
From pipe provers to small volume provers
Many petrochemical installations have a permanent pipe prover, a volume displacement prover that uses a sphere as the displacer instead of a piston. Pipe provers need a relatively large pipe segment to give accurate results, so they are large installations, not suited to mobile use. As technology developed, small volume provers (SVPs) became a viable alternative. They are smaller than pipe provers, so they can be used in a mobile setup, which makes them suitable for local weights and measures offices.
In February 2023, staff of NIST's Office of Weights and Measures (OWM) were invited to watch the Michigan Department of Agriculture's E.C. Heffron state metrology laboratory calibrate an SVP. They wanted to learn about the calibration procedures used in state metrology laboratories and the critical factors in them. The Michigan lab calibrates about 15 SVPs a year, with capacities from 5 to 30 gallons, mostly for the petrochemical industry.
How an SVP measures
A volume displacement prover has three parts: a pipe of known capacity, a device that pushes fluid through it, and sensors that detect when the fluid reaches the start and the end of its travel. In an SVP the displacer is a piston on a shaft that ends outside the enclosure. A trigger on the shaft sets off two sensors on a neighboring bar as the piston moves (A and B in the first figure), and the distance between the sensors defines the calibrated volume inside the enclosure. The exact volume between the sensors is verified in a laboratory calibration.

Schematic of a Small Volume Prover. Figure from NIST's article.
The prover is installed in line with the meter under test, which produces a pulse for every unit of volume that passes through it. When the displacer passes point A, the measurement system starts counting the meter's pulses; when it passes point B, the count stops. The meter and the prover do not hold exactly the same liquid at the same moment, but the quantities match: the amount that traveled from A to B equals the amount that passed through the meter in the same time. Since the volume between A and B is known, it should match the number of pulses multiplied by the volume per pulse.
The piston has a seal so that no liquid slips past it while it crosses the calibrated volume. It also has a poppet valve that opens when the piston is pulled back to its starting position, so the flow of liquid is not interrupted.
Calibrating an SVP in Michigan
The Michigan lab follows NIST Standard Operating Procedure 26. The aim is to calibrate the internal volume between the two trigger points, then compare it with the volume found at the last calibration to detect any drift that could point to a defect in the prover. The tolerance applied for repeatability is 0.02 percent.
The lab uses demineralized water from a 2,000-gallon (7500-liter) tank, which keeps the pressure sufficiently constant. The prover is filled with water and all air is vented. As water flows it slowly pushes the piston forward. Until the trigger reaches the first sensor the water coming out is drained; between the first and second sensors it is automatically diverted into a collection tank; after the second sensor it goes back to the drain. With a constant flow, the water collected equals the water held in the calibrated volume.
The lab weighs the tank before and after on a 600 kg x 0.1 g scale used as a mass comparator, and the difference is the weight of the water. That weight is converted to a volume with the calculations in SOP 26, with corrections for the temperature and pressure of the water inside the prover. A calibrated pressure sensor sits on an access point on the prover and a calibrated temperature sensor in a thermowell in the enclosure.
Ideally the calibration runs at the flow rate the prover sees in the field, but many state laboratories cannot manage that. A 20-gallon prover like the one calibrated during the visit normally operates at about 500 gallons per minute, while the maximum the Michigan setup can reach is about 3 gallons per minute. To check the piston seals, the lab runs the calibration at the high rate (about 3 gallons per minute) and at a low rate (about 2 gallons per minute), five times in the order high, low, high, low, high. The difference between the high and low results is examined closely, because any variation often means fluid is seeping past the piston seals. If the calibrated volume has changed significantly, the new value is entered in the control software of the prover's measuring system.
What affects accuracy
An SVP can detect a predetermined amount of fluid displacement accurately, provided three influences are managed:
| Factor | What can go wrong | What is done |
|---|---|---|
| Controlled flow | Liquid seeps past the piston or through the poppet valve. Both seals wear. | Seals are replaced periodically. Some provers detect leaks in the piston seals; others record the number of operations as a gauge of seal age, so replacement can be planned. |
| Flow consistency | The prover and the meter sit at different points in the pipeline, so they do not measure the identical batch of fluid. Air bubbles, contamination or an uneven mixture upset the comparison. | SVPs are used only where the composition of the product is relatively constant. |
| Temperature | A warmer prover expands, and its calibrated volume grows. The sensor bar, outside the fluid compartment and more exposed to the surroundings, also expands, which moves the sensors apart and again enlarges the volume. | The prover's software compensates for both the fluid temperature and the sensor bar temperature during a meter calibration. |
Sources
- National Institute of Standards and Technology, Office of Weights and Measures, Small Volume Provers (SVPs): Small-Scale Measurement Devices with a Big Impact, which cites the U.S. Energy Information Administration's frequently asked questions for the petroleum figures; rewritten in hubnx's own words.