A scanning electron microscope (SEM) that dirties the sample it is measuring is measuring the dirt. Since the SEM's inception contamination has dogged its users, and at the nanometer scale it becomes a showstopper. NIST's Physical Measurement Laboratory (PML) has spent years working out how to remove it. A NIST news article tells the story.

What contamination does

Under a focused beam of electrons or ions, a carbon-based film gradually builds up on the sample, leaving characteristic dark patterns. It is typically caused by residual organic material inside the SEM itself. The buildup changes the sample and changes the number, path and energy of the electrons that leave it, so repeatable, quantitative measurements become difficult or impossible. The stakes are high in the integrated circuit industry, where a 1 nm improvement is estimated to be worth 1 billion dollars or more in added economic value.

Gold-on-carbon sample before and after cleaning: heavy beam-induced contamination after 10 minutes of imaging, and a clean image after hydrogen plasma cleaning

An evaporated gold-on-carbon sample imaged first, then (b) after 10 minutes of continuous imaging at twice the magnification, with excessive electron-beam contamination, and (c) after hydrogen plasma cleaning and the same 10 minutes of imaging, without it. NIST.

The long fight

The battle began in the 1990s, before András Vladár joined NIST, led by Michael Postek of the Semiconductor and Dimensional Metrology Division. Working with a small U.S. company, XEI Scientific, they tested a prototype low-energy (20 W) plasma unit that attaches to an SEM access port and cleans the instrument by flooding it with oxygen plasma.

When Vladár joined the SEM project in 1999, research continued: the best plasma pressure range, how long cleaning really needs. The team realized that a clean instrument is not enough; a clean reference sample is essential too. With a known-clean sample, you can tell instrument dirt from sample dirt.

The sample cleaning apparatus built at NIST, with the plasma cleaning unit attached, and a close-up of a sample being cleaned

The NIST-built apparatus for cleaning samples, with the low-energy plasma unit attached at the right, and a sample being cleaned. NIST.

A second collaboration, with IBSS Group, produced a modified plasma cleaner that works over a wider vacuum range and at higher power, up to 99 watts. Low power is enough for oxygen plasma, but samples sensitive to oxygen, such as evaporated gold on carbon, need hydrogen or helium plasma, and there higher power helps.

Results

The decades of work produced a NIST Contamination Specification: steps SEM users worldwide can follow to find out whether an instrument needs cleaning and how to clean it effectively. More than 2,300 low-energy plasma-cleaning devices are now in use. According to Vladár, a cleaned instrument can stay free of contamination for months unless a dirty sample is put in, and the work shows that government researchers teaming with small U.S. companies can help solve serious problems in research and industry.

What may come next

New PML research suggests that the electron bombardment during long measurements can itself clean many kinds of samples. Vladár hypothesizes that any surface exposed to air picks up a very thin layer of water, and that the electron beam may remove it. If that holds, SEM users might reach ultimate cleanliness without an ultra-high vacuum, which could make measurements easier and cheaper.

Hydrogen-cleaned gold-on-carbon sample, magnified, imaged initially and after 10 minutes of imaging at twice the magnification, with better resolution and more detail

The magnified center of a hydrogen-cleaned gold-on-carbon sample, imaged initially (left) and after 10 minutes at twice the magnification (right): better resolution and more detail after both hydrogen plasma and electron cleaning. NIST.

Next the lab wants to learn exactly how and why the cleaning works. Meanwhile the method has spread to other tools, including transmission electron microscopes and extreme ultraviolet imagers.

Sources

Комментарии

Spinner Logo

Ещё на тему «Наука»

Автор: Hub Nexus editors
Top 10 in AI, week of 19 October 2026
Автор: Hub Nexus editors
Top 10 in AI, week of 12 October 2026
Автор: Hub Nexus editors
Top 10 in AI, week of 2 November 2026
Автор: Hub Nexus editors
Top 10 in AI, week of 26 October 2026
CC0 1.0 (общественное достояние)
Autism prevalence among 8-year-olds was 1 in 59 in 2014. The figure most often missed: 44% of the children identified had an IQ above 85. And while 85% had a developmental concern recorded by 36 months, only 42% had been comprehensively evaluated by then.
CC0 1.0 (общественное достояние)
More than 70 percent of U.S. adults are overweight or have obesity, raising the risk of diabetes, heart and kidney disease and some cancers. What drives weight gain, and what helps.