For almost 400 years mercury gauges have been the most accurate way to measure pressure. Within weeks of its "first light," a new National Institute of Standards and Technology (NIST) device surpassed the best mercury-based techniques in resolution, speed and range, in an instrument far smaller. It is called a fixed-length optical cavity (FLOC), and it works by detecting tiny changes in the wavelength of light passing through a cavity filled with nitrogen gas.

The FLOC inside its copper container with the cover removed, revealing the upper of the two channels.

The FLOC inside its copper container with the cover removed, revealing the upper of the two channels. Image from NIST's page.

How it compares

The team, from NIST's Thermodynamic Metrology and Dimensional Metrology groups, expects the FLOC to replace traditional mercury manometers as the standard used to calibrate commercial equipment. It is about the size of a travel mug. The research group leader says it is the first photonic-based primary pressure standard, and it is a promising basis for a factory-floor instrument for semiconductor, glass and aerospace manufacturing.

FLOCNIST's mercury standard
SizeAbout the size of a travel mugA 3-meter (about 10-foot) column of mercury through the calibration room ceiling
Resolution0.1 mPa (millipascal)36 times coarser
SpeedAbout a secondAbout a minute and a half, 100 times slower
RangeMillipascals up to the hundred thousand pascal level, wider at the low endNarrower at the low end

How it works

Mercury instruments read pressure as the height of a column. They are accurate and reliable but bulky, and mercury is neurotoxic, so a global effort to phase it out has produced a United Nations Environment Program treaty signed by more than 100 countries.

The FLOC is a temperature-controlled optical cavity about 15.5 cm long and 5 cm by 5 cm, encased in copper, with two channels: one filled with nitrogen gas and one in vacuum. A low-power red laser beam (633 nm) is locked to each channel, forming a standing wave that keeps adjusting its wavelength. Some of the light from each channel leaves the device and the two beams combine into an interference pattern. Pressure does not affect the light in the vacuum, but it changes the density of the nitrogen, and so how fast light travels in it, which shifts the wavelength of the light in that channel. The shifts are on the order of picometers, a hundred times smaller than an atom, and show up in the interference pattern.

FLOC set-up

The FLOC setup: laser-directing optics (right), copper-enclosed cavity (center) and the output signal on a computer screen (left). Image from NIST's page.

What is still to do

So far the FLOC is accurate to within 0.005%, or 50 parts per million (ppm), better than most commercial pressure instruments but not yet better than the mercury manometer. Over the next three years the researchers want to push the uncertainty below 5 ppm, which would surpass the manometer standard. They also see the system shrinking, perhaps to the size of a smartphone, while becoming faster and finer.

Sources

Комментарии

Spinner Logo

Ещё на тему «Наука»

Автор: Hub Nexus editors
Top 10 in AI, week of 26 October 2026
Автор: Hub Nexus editors
Top 10 in AI, week of 19 October 2026
Автор: Hub Nexus editors
Top 10 in AI, week of 12 October 2026
Автор: Hub Nexus editors
Top 10 in AI, week of 2 November 2026
CC0 1.0 (общественное достояние)
Alaska's Deep-Sea Sponges: Building a Library of Species Nobody Has Named Yet
CC0 1.0 (общественное достояние)
Dall’s Porpoise Expands Territory in a Changing Prince William Sound