Microwaves with wavelengths of a centimeter down to a few millimeters can provide nanometer-scale resolution. That sounds like a paradox, and it is the basis of the near-field scanning microwave microscope (NSMM), a new version of which was about to arrive at NIST, according to NIST. NIST's ability to determine the composition and physics of nanoscale materials and devices was about to improve dramatically.

The larger of the two alternating images shows the topography of a nanowire bundle. The smaller image shows the same bundle without the p-doped portion; the difference lets researchers estimate where the p-n junction lies.
What an NSMM does
Few techniques can measure at this resolution across so wide a range of samples: semiconductors, semiconducting nanowire, photovoltaic materials, magnetic and multiferroic materials, and even proteins and DNA. Researchers in the Electromagnetics Division of NIST's Physical Measurement Laboratory (PML), using commercial and home-made NSMMs, pioneered many applications, notably mapping semiconductor dopant distribution (charge-carrier concentration) in 2D and 3D. Emerging technologies such as molecular electronics, carbon nanotubes, nanowires, graphene and spin-based electronics need that ability.
Mitch Wallis of the Radio-Frequency Electronics Group explained the motivation: combine the fine spatial resolution of scanning probe instruments with the broadband reach of microwave measurements, to study magnetic and mechanical resonance at the nanoscale. Phones and computers work at a few gigahertz, so the nanoscale parts inside them must be measured at those frequencies to learn how they perform.
How it works

Schematic diagram of NSMM apparatus.
An NSMM is an atomic force microscope (AFM) with a continuous or pulsed microwave signal applied to its tip. The tip scans in soft contact or a few nanometers above the surface and emits a microwave signal that the material scatters, changing its frequency, amplitude and other properties. The tip also serves as the receiving antenna. The returned signal depends on variables such as permeability, permittivity, sheet resistance, dielectric constant, impedance and how capacitance changes with voltage, all set by the sample's composition.
At each point in an area, typically a few micrometers square, the tip's position is merged with the returned signal to build a composite image. NSMMs can be tuned to image from sub-micrometer to 100 μm below the surface, so a thin metal layer over a piezoelectric material can be seen through.
Why does it work? Pavel Kabos of the Advanced High-Frequency Devices Program says the AFM tip is only about 100 nanometers wide and sits a few nanometers from the surface or in soft contact, and that this tip elevation governs the resolution. Because that distance is far smaller than the wavelength, near-field effects predominate, allowing nanometer resolution and accurate quantitative measurements.
Two ways to probe

PML's home-made NSMM, with the tip poised above a 2 cm wide sample.
- Change the tip voltage. In a 2012 paper, Atif Imtiaz and PML colleagues, with researchers at Agilent Technologies, showed that the sign of the bias attracts or depletes charge carriers at the surface, forming a space charge region whose width depends on carrier density. That gives a sensitive measure of dopant composition. Kabos said they had recently looked into a 3 micrometer wide area to see how doping is distributed at a P-N junction and locate it precisely.
- Change the microwave frequency. The same material may look very different at 5, 7 or 18 GHz, so researchers can find the local doping profile and also performance at a given frequency in a given region. Chip designers want such doping profiles around transistor gates, sources and drains in memory chips.
The new instrument and what is next
The existing instruments have one tip, open to the air. The new NSMM, slated to arrive that fall, has four tips for simultaneous comparison of materials and sits in an ultra-high vacuum chamber to reduce interference and contamination. It also has a vacuum sample-preparation chamber and temperature control from about 30 K to room temperature. It will allow closer study of materials for nanoscale electronics, bioelectronics and spintronic or phasetronic applications.
Planned work includes:
- reference materials, with reliable quantitative models for local properties;
- combining voltage- and frequency-dependent spectroscopy, such as measuring the derivative of the phase between incident and reflected signals alongside the derivative of capacitance with voltage, which could improve signal-to-noise, sensitivity to water in biological samples and perhaps imaging of living cells in saline;
- two-probe and multi-probe NSMM setups.