In the United States about 80 million CT scans are made every year, 7 million of them on children, according to the American Association of Physicists in Medicine (AAPM). Hence the intense interest in measuring exactly what dose patients receive. NIST's Heather Chen-Mayer and colleagues from the Radioactivity and Dosimetry Groups have shown that measuring the temperature change that the dose causes in plastic phantoms, stand-ins for human tissue, is a valuable complement to the conventional method and might eventually serve as a primary standard.
The old way and its limits
CT dose is now assessed with the computed tomography dose index (CTDI), introduced more than 30 years ago, developed under AAPM guidance and adopted by the FDA as a regulatory performance standard. It needs a 10 cm ionization chamber placed in a standard cylindrical phantom made of polymethyl methacrylate (PMMA). X-rays ionize gas in the chamber and the charge collected is converted to dose with a calibration coefficient from in-air measurements, linked to NIST's air kerma standard. Modern CT, with wider beams and more complex protocols, has put CTDI under scrutiny. "Many experts regard it as inadequate," says Chen-Mayer, since it is not ideally suited to the helical scan pattern. AAPM task groups are working on phantom design and measurement practice, and on showing how dose depends on phantom size.
NIST built a version of the proposed AAPM phantom from three sections of high-density polyethylene (HDPE) cylinders and joined measurements with ionization chambers.

Sensor placement in the phantom: in the center channel an x-ray image of an ionization chamber (left) and thermistors (right).
Try calorimetry
In February 2012 the team decided to try calorimetry, which NIST has used for decades. It measures the tiny temperature change as x-rays deposit energy in a material, and is a primary reference standard for therapy-level dose. As far as Chen-Mayer knows, no one else was doing it in CT.
HDPE's density (0.93 to 0.97 g/cm³) is closer to water's (nominally 1.0), the usual tissue surrogate, than PMMA's (about 1.2 g/cm³); for a typical CT spectrum in a phantom 20 cm or wider, dose in HDPE resembles dose in water. It also conducts heat less than water, allowing longer measurements.
The worry was signal size: CT doses are far below therapy doses.
| Radiation | Temperature rise |
|---|---|
| 1 Gy in water | 0.24 mK |
| 1 Gy in HDPE | 0.6 mK |
| Typical CT scan (a few mGy) in HDPE | in the µK range, close to calorimetry's detection threshold |
The team first tried a 6 MV medical linac beam, with doses of a few grays, reasoning that if it failed there it would fail at CT's milligray level. It worked and agreed with ionization chambers and water calorimeters. They then moved the phantom to PML's new PET/CT machine and took readings from 20 consecutive 2.8 s axial scans to get a sizable signal.
The heat defect
Readings must be corrected for HDPE's thermal conductivity, which is straightforward. Harder are photolytic chemical reactions in the plastic that can add or absorb heat, a "heat defect", says team member Ron Tosh, also seen in water calorimetry because of dissolved gases. The researchers believe it could change values by as much as 10%. They are now testing thermistors embedded in polystyrene, whose heat defect is zero.

From bottom left: an ionization chamber in its casing, a polystyrene enclosure for thermistors (to be implemented) and two thermistors. The four black items at right are thermistor covers.
A standard calorimeter for CT's divergent, rotating fields will take much work, but the team has established the potential of calorimetry for CT dosimetry and the possibility that it becomes a primary standard. The proposed 30 cm HDPE phantom would replace the current CTDI phantoms (32 cm for the body, 16 cm for the head); the study used 30 cm, 26 cm and 13 cm.