Proteins in a living body are constantly in motion: they bend and flex in different ways at different points, and each molecule's vibrational dynamics affect its propensity to bind to neighbors such as drugs. Terahertz (THz) spectroscopy, a newly emergent field using frequencies between microwaves and infrared radiation, examines biosystems by their vibrational response. Researchers in NIST's Physical Measurement Laboratory (PML), in the Sensor Science and Radiation and Biomolecular Physics divisions, are providing baseline data for it with high-resolution instruments.

Why THz

A biomolecule's vibrational modes depend on many factors: conformational variety, hydration state, solvate environment, peptide sequence, crystalline order and more. Academic labs generally use low-resolution spectrometers and often lack computational chemistry expertise, which limits their results. Zeeshan Ahmed of the Temperature & Humidity Group says that sources and detectors for this range did not exist until recently, and it used to be called the "terahertz gap"; telecommunications advances opened it up.

Other methods fall short for this question. X-ray crystallography gives a sharp picture of a single conformation, while proteins in the body keep moving. Infrared modes, says David Plusquellic of the Biophysics Group, are usually localized to specific regions, whereas THz modes involve the collective motions of all atoms in the structure.

How a measurement works

Researchers scan an isolated sample with frequencies typically in the range from 0.06 THz to 10 THz. Frequencies that match a vibrational mode are absorbed and the rest pass through to a detector, giving absorption spectra that can reveal differences between molecules that look virtually identical in IR.

At PML the process has several stages:

  1. Study the molecule's x-ray and IR spectra and calculate the vibrations predicted by quantum chemistry.
  2. Scan a sample in vacuum at room temperature, at 77 K and at 4.8 K; the coldest gives the sharpest absorption features.
  3. Prepare the sample either as a pellet, mixed with polyethylene (transparent in the THz range) and compressed into a disk about 1 cm wide and 1 mm to 3 mm deep, or as a thin film in a waveguide.

Sample for thin-film measurement

In the waveguide method, the thin-film sample sits between two gold-coated plates about 250 µm apart. It gives much higher line resolution than the pellet.

Graphs of THz spectra

Left: THz spectroscopy (blue lines) clearly distinguishes three solid forms of the polypeptide trialanine, while commonly used infrared (black lines) only separates parallel (p-AAA) and antiparallel (ap-AAA) forms, not hydration levels. Right: better resolution in a waveguide (blue lines) than as a pellet (red lines).

Water and drugs

THz spectroscopy holds promise for studying how biomolecules' motions change when bound to water. Ahmed, Plusquellic and colleagues showed that water changes the structure and vibrational modes of three peptides common in protein systems, and that THz distinguishes the variations clearly. They also found the energetically preferred states of crystal hydration and suggested ways to reconcile theory with observed spectra. Sensitivity to solid forms and water has applications in process engineering, crystal engineering, IP patrolling, quality control and spotting counterfeit drugs.

PML has extended the method to pharmaceuticals. Ahmed notes that dehydrated drugs are generally absorbed faster in the body, so hydration state can change a patient's drug exposure and the dosing profile the FDA approved, and the solid form must stay the same over a production campaign spanning many years and during storage. Data now exist for non-steroidal anti-inflammatory drugs such as Naproxen and Norfloxacin and antibiotics such as Amoxicillin and Ampicillin.

Challenges and aims

Plusquellic says major challenges remain: often the researchers are unsure what spectra are telling them, and theoretical calculations often disagree completely with observation. One of PML's most important contributions is sets of experimental data against which theory can be tested. Ahmed would like to characterize a drug bound to a peptide or protein, since x-ray crystallography does not reveal hydrogen bonding but THz vibrations do. The long-term aim is better theory, so that many steps in drug discovery can be done by computer, which could reduce research costs dramatically and help the industry bend the "Eroom's Law" curve: the decline in new drugs approved per unit of development cost despite technological advances, R&D support and regulatory changes. The PML work supplies baseline benchmarks to improve the accuracy of computational methods and help researchers choose the one best suited to their needs.

Sources

Комментарии

Spinner Logo

Ещё на тему «Наука»

Автор: Hub Nexus editors
Top 10 in AI, week of 12 October 2026
Автор: Hub Nexus editors
Top 10 in AI, week of 2 November 2026
Автор: Hub Nexus editors
Top 10 in AI, week of 26 October 2026
Автор: Hub Nexus editors
Top 10 in AI, week of 19 October 2026
CC0 1.0 (общественное достояние)
More than 70 percent of U.S. adults are overweight or have obesity, raising the risk of diabetes, heart and kidney disease and some cancers. What drives weight gain, and what helps.
CC0 1.0 (общественное достояние)
What Ohio gets from being mapped by laser