Автора пока нетВзять себе
- Лицензия
- CC0 1.0 (общественное достояние)По материалам www.nist.gov
- Языки
- English
- Темы
- Наука
When the kilogram is redefined, the new definition will be realized in vacuum, but the weights people actually use are weighed in air. Bridging that gap is hard, and a group at the National Institute of Standards and Technology (NIST) proposed a way around it, as this 2012 article describes: levitate the mass with magnets.
The problem
The planned methods of realizing the redefined kilogram are based on the Planck constant and the Avogadro constant, and both work in vacuum. Secondary standards, and the transfer standards used in laboratories and industry, must work in air. Two masses are easy to compare when both are in vacuum, but in air a measured mass is affected by air buoyancy and by water and other contaminants adsorbed on its surface. Correcting for adsorbed water needs the mass's volume and the air's pressure, temperature and humidity, fed into a model that accounts for the surface properties of the metal. There is no firm consensus data to base those models on; measurements of water adsorption onto stainless steel have recorded mass changes that differ by as much as a factor of 15.
NIST's alternative
The Quantum Measurement Division of NIST's Physical Measurement Laboratory (PML) designed a system for direct comparison of a reference mass in vacuum with an unknown artifact in air, using magnetic levitation. The Mass and Force Group showed proof of principle with a small tabletop model and was building a full-scale apparatus about three meters tall. Patrick Abbott of the Mass and Force Group said most of the world's effort was going into adsorption models, and that he expected others to adopt a direct comparison that depends on no particular material or adsorption assumption once it was shown to work.
How the apparatus works
The full-scale device is built of aluminum, with two adjacent enclosures.
- Vacuum chamber (top). It holds the reference mass and a specially modified mass comparator that can resolve 10 micrograms in a kilogram, 10 parts in 10^9. The comparator's weighing pan can couple directly to the reference mass or to a solenoid-and-permanent-magnet combination hanging just above the chamber floor.
- Air chamber (below, behind an airtight barrier). It holds the artifact to be calibrated, on a suspension carriage topped by a permanent magnet with one exposed pole. That pole is attracted to the opposite pole of the magnet above, and when the poles are the right distance apart, typically 15 to 17 mm, the artifact and carriage levitate.
- Feedback. A computer controller adjusts the current in the solenoid's roughly 2500 windings 20,000 times a second to hold equilibrium. The first design used a laser and receptor to watch the mass's position; the researchers then switched to a Hall-effect sensor on a cantilever that tracks the magnetic field between the poles, which depends on their separation. The magnets are shielded so stray flux does not disturb the rest of the system.
If the magnetic force exceeds the weight, the poles snap together; if it is weaker, the mass sinks. For each load and pole distance there is one field strength that gives near-motionless levitation.
Magnetic levitation is not new, since objects from passenger trains to live frogs have been levitated, and small commercial systems have weighed up to 200 grams this way. What is new, Abbott said, is weighing up to several kilograms with higher precision than before, moving between vacuum and air, and using levitation to pass on mass values.
Where it stood
A 50 cm tall working model, able to measure loads from about 100 g to a little over a kilogram, was used for the proof of principle. Its stability, limited by the balance's resolution, was about 1 mg in 1 kg, or one part in 10^6. The project's target is an uncertainty of no more than a few parts in 10^8, only slightly above what the international Consultative Committee for Mass and Related Quantities has recommended for realizing the kilogram for primary standards. The researchers hoped to have initial results from the full-scale apparatus around December 2012.
Though the device avoids adsorption modeling, it is also expected to help it: the group planned to make direct measurements and compare them with the results of people using adsorption models, as verification for both sides. The Mass and Force Group is led by Zeina Kubarych.
Sources
- National Institute of Standards and Technology, Disseminating the Kilogram, No Strings Attached; quotations from NIST staff are paraphrased; rewritten in hubnx's own words. The article's two photographs of a named researcher are left out.