Автора пока нетВзять себе
- Лицензия
- CC0 1.0 (общественное достояние)По материалам www.nist.gov
- Языки
- English
- Темы
- Наука
Redefining the kilogram by a fixed physical constant will not end the headaches over mass that drifts. When the story was written, such a redefinition was expected soon, but in practice mass would still be passed along as physical artifacts, and those would still be open to environmental effects such as gases sticking to their surfaces or leaving them.
That is why Joshua Pomeroy and his colleagues in the Mass and Force Group, part of the Quantum Measurement Division of NIST's Physical Measurement Laboratory, built and are now testing an experimental apparatus that studies how different material surfaces gain or lose mass in different environments.
Why artifacts drift
Many experts hope for an "infallible" artifact whose material or coating would make it virtually impervious to mass variation. The platinum-iridium International Prototype Kilogram is the kind of artifact whose mass varies over time. Pomeroy says such a perpetually stable object may not be possible. NIST therefore wants to understand how surface mass changes with time and to build models that capture it, which it can give to customers so they can correct their measurements.
The instrument

The double-paddle oscillator (DPO), gray, mounted face down before it goes into the vacuum chamber. An increase (decrease) in its resonant frequency indicates a loss (gain) of mass.
| Feature | Detail |
|---|---|
| Sensitivity | 0.02 nanograms per square centimeter, equivalent to the mass of a penny spread over an area of 11 km2 |
| Compared with the best commercial device | About twice the sensitivity of a quartz crystal microbalance (QCM) |
| The oscillator | A thin silicon crystal cut into two paddle-shaped sections joined by a central axis; one vibrational mode has a very high quality factor and a sharply defined resonance frequency |
| Surface area | 6.65 cm2, about 1 square inch and about 13 times larger than the surface of a QCM |
| Chamber | Vacuum, with radiation shielding; temperature adjustable from 500 K to a few K |
| Deposition source | A four-pocket vapor deposition source beneath the face-down DPO can place as many as four different substances on it, including by co-deposition |
The goal is to imitate the surfaces of possible mass-artifact materials at different temperatures and in different gas environments, and to gather quantitative, repeatable measurements of what happens to artifacts in situations such as transit between Beijing and Mexico City.

(a) Test chamber with the cryogenic cooler at top, vacuum pump at bottom and residual gas analyzer at left. (b) Geometry and dimensions of the double-paddle oscillator. (c) Close-up of the sample mount with a DPO assembled. Image from NIST's page.
What comes first
Deposition experiments have begun with nitrogen, a good first test gas because it is safe, makes up 78% of air, and at 10 K has a sticking coefficient close to 1, meaning essentially all the molecules put into the chamber stick to the DPO. It will be months before the device is fully characterized. Only then can comparison studies of candidate materials and coatings for a new class of mass standards begin.
Pomeroy cautions against hoping for a silver bullet. Some contributors to mass variation, such as microscratches, cannot be controlled, and even good models of simple chemical adsorption will have to account for effects such as porosity. Diamond-like carbon coatings have been suggested and may prove the material of choice, but they can have irregular pinholes and voids that change porosity and adsorption.
The timing question
Periodic measurements of the platinum-iridium prototypes show excursions over 100 years, but whether the change is gradual or abrupt is not fully understood. The International Bureau of Weights and Measures (BIPM) sets a cleaning protocol before a standard is recalibrated. Pomeroy describes the pattern: the mass goes down after cleaning, then rises again, and after some time the standard is calibrated. A "growth" model developed by the BIPM and the calibrated value are used to predict the mass at a future date. Pomeroy says little is known about changes on shorter timescales: the prototypes are known to drift over decades, but nobody knows when the change happens. It could be in the first five minutes after cleaning.
Pomeroy also plans to build a research-grade QCM into the instrument and compare it with the DPO. QCM makers and users sometimes ask the group for NIST traceability, and there is currently no system for it, so this work could open a path to offering it.
Pomeroy compares the effort to building a flashlight to shine around a cave: the miners do not yet know which ores are there or which are worth extracting, but they need the flashlight to find out.
Sources
- National Institute of Standards and Technology, "Surface Detail: Oscillator to Explore Secrets of Mass Variation" (https://www.nist.gov/news-events/news/2014/01/surface-detail-oscillator-explore-secrets-mass-variation); a work of the United States government; rewritten in hubnx's own words.
Комментарии






