LED lamps stay on the market only about 12 months, according to NIST’s Optical Radiation Group, so the program that checks a laboratory’s measurements of them has to be renewed regularly. NIST now offers the second generation of its Solid-State Lighting Measurement Assurance Program, MAP-2, with a new set of lamp artifacts, detailed measurement protocols and revised operations. Demand for solid-state lighting is rising in homes, businesses and elsewhere, and manufacturers and test labs need to measure total lumen output, color temperature, hue and saturation, power use and efficiency accurately.
How a lab takes part
- NIST sends a kit of lamps whose properties it has measured extensively and precisely.
- The customer measures them following NIST’s protocols and sends the results back along with the lamps.
- NIST re-measures each lamp in case shipping affected it, averages the values with the earlier ones and returns a report comparing the customer’s results with NIST’s.
In general, a difference of more than plus or minus 4% is scored “fail” for that measurement. A participant that fails one or more measurements takes corrective action. If that means retesting a lamp, the lab must contact NIST and arrange, for a fee, to receive a similar lamp. According to NIST, nearly all labs eventually correct the problems.

A wider view of the laboratory area. Image from NIST’s page.
What is in MAP-2
- It began in January 2015.
- An updated set of lamps for general illumination represents the current residential market, with some chosen for specific measurement challenges.
- Participants may also receive, free, two of the latest 4-foot LED tube lights: one that works with a conventional ballast like a fluorescent tube fixture, and one that runs on 120 volt, 60 Hz current without a ballast.
- Each participant is asked to measure the output of an included incandescent bulb on both alternating and direct current. The results are not part of the pass/fail analysis, but will add to NIST’s understanding of problems with 4-pole sockets, since the lamp draws a lot of current.
Who it is for
The program matters most to laboratories that need periodic accreditation and proficiency testing, typically every two or three years, for example to meet the Illuminating Engineering Society standard IES LM-79 (developed with support from PML staff) or to take part in the Department of Energy’s Energy Star program. MAP-1 ran for four years and served 120 customers worldwide. MAP-2 is expected to run for three years and is open to any testing laboratory. The fee ranges from $4000.00 for a kit for 8 or 10 measurements to $5400.00 for lamps for 12 different measurements, including some on compact fluorescent lights.
A snapshot of the field
NIST collects data anonymously from all customers. The leader of the Optical Radiation Group said that gives a snapshot of the state of the measurement community, showing where standards need to improve and helping to identify factors that are still poorly understood.