A pesticide made by bacteria is a mixture of related molecules, and its makers must show they can measure how much of the active ingredient is in each bottle. A U.S. Geological Survey report describes how its scientists did that for antimycin-A (ANT-A), a chemical used for decades to remove unwanted fish.
Why a method was needed
Fisheries and aquaculture managers have often used pesticide formulations with active ingredients such as rotenone, ANT-A and TFM to remove nonindigenous fish. ANT-A use fell for several years after its pesticide registration lapsed. The USGS Upper Midwest Environmental Sciences Center (UMESC) has since obtained the right and possession of a strain of Streptomyces bacteria that yields a lot of ANT-A, and is preparing a dossier to reinstate a liquid ANT-A formulation with the Environmental Protection Agency (EPA). Reinstatement needs a validated way to measure the active ingredient in the liquid, as federal pesticide law requires.
The difficulty is that ANT-A, as a product of fermentation, is several related compounds (homologs and stereoisomers) whose proportions vary between batches and manufacturers. Four homologs, A1 to A4, usually make up 80 to 100 percent of a batch's purity. Early checks of potency used bioassays on fish and yeast. Later liquid chromatography-mass spectrometry (LC-MS) work could measure ANT-A in water at nanograms per liter, but one earlier method, which calibrates each homolog separately and adds them up, did not always meet the usual requirement that a calibration curve have a coefficient of determination (r²) of at least 0.995.

Chemical structure of antimycin-A; the two side groups, R1 and R2, differ among homologs (USGS).
The new approach
Instead of one calibration curve for each homolog, the USGS team built a single curve from the combined, averaged response of the four main homologs. This copes better when a standard or sample has some homologs in low abundance, and it holds up when the mix changes from batch to batch.
- Standards. Six calibration standards covered 1 to 450 micrograms per liter.
- Samples. Two solutions were made from solid ANT-A: solution 1 dissolved in the solvent of the old registered formulation, and solution 2 also combined with the formulation's inert ingredients. Solution 2 was cleaned up on a solid-phase extraction cartridge to remove non-target compounds, with ANT-A eluted in acetonitrile rather than the acetone used before, which improved stability and accuracy.
- Instrument. An HPLC separated the homologs and a triple quadrupole mass spectrometer identified them by retention time, precursor ions and product ions. One product ion, 245.0 m/z, served as the quantifier for all four.

The four homologs' peaks from one injection of a 100 micrograms per liter sample. They elute between 3.3 and 3.9 minutes, in the order A4, A3, A2, A1 (USGS).
How well it worked
The criteria were a curve r² of at least 0.995, accuracy within plus or minus 20 percent, and precision under 20 percent relative standard deviation. Every calibration curve met the r² criterion. Three analyst groups ran the samples.
| Solution 1 (nominal 230 µg/L) | Solution 2 (nominal 100 µg/L) | |
|---|---|---|
| Group accuracy | 97.8, 98.5 and 98.8 percent | 114.9, 96.7 and 84.4 percent |
| All three combined | 98.4 percent (± 1.4) | 98.7 percent (± 5.7) |
| Precision across groups (relative standard deviation) | 2.5 to 5.7 percent | 4.8 to 16.7 percent |
Differences between the groups were not statistically significant for either solution. The authors conclude that ANT-A can be measured in a liquid formulation with high accuracy (98 to 99 percent), precision and specificity, reproducibly, and that the method is ready for consideration in regulatory use.
What is next
The method uses multiple reaction monitoring instead of the single ion monitoring used before, which adds selectivity while keeping similar sensitivity, and an advanced column and modified solvents shorten run times. It was designed for the formulation itself, not for water: the earlier method's limits are low enough for surface-water monitoring. Future work could test whether this approach and earlier water-sampling methods can measure a liquid formulation in water during native-fish restoration projects.
Sources
- U.S. Geological Survey, Determination of antimycin-a in a liquid formulation by high performance liquid chromatography-mass spectrometry; rewritten in hubnx's own words and condensed.