- Licence
- CC BY-SA 4.0Adapted from www.appropedia.org
- Languages
- English
A paper in the Journal of Water, Sanitation and Hygiene for Development describes a water testing platform whose electronics are open-source and whose case is 3D printed, and which its authors say is as accurate as commercial equipment while costing between 7.5 and 15 times less. It grows out of an open-source colorimeter and belongs to a larger effort to cut the price of scientific equipment with open-source hardware.
Why it was built
Many people in the developing world lack safe drinking water. Cheap ways of treating water exist, but deciding whether treatment is needed, whether it works and whether it stays reliable requires cheap testing too. The paper offers a method for designing, building and technically validating a low-cost open-source tester, with a case study that measures chemical oxygen demand (COD) and biochemical oxygen demand by colorimetry, plus turbidity by nephelometry under ISO 7027. The authors conclude that open-source hardware is a promising way to get the equipment needed for water quality measurement in developed and developing regions alike.
What goes into it
- an Arduino Fio, with a programmer board if needed;
- an Adafruit LCD shield and a 3.3 V LCD (a 5 V screen needs an external power supply);
- one 860 nm infrared LED and one 620 nm orange LED;
- three TSL235R light-to-frequency sensors;
- wires, solder and filament for the printed parts.
Building it means soldering the LCD shield following Adafruit's tutorial, wiring the LEDs, screen and sensors to the Arduino, uploading the firmware, which is on GitHub, and testing. Each LED is paired with a reference sensor, and a third sensor sits on its own. The firmware needs two extra libraries, LcdMenu and DuplexFrequencyCounter. If the printed case warps, the page suggests printing it in two parts: the sample chamber with the sensors, and the box for the screen and Arduino.
Using it
The menu offers two tests, COD and nephelometry, in the order they are normally used.
| Test | Steps |
|---|---|
| COD | Zero (cancels the orange LED's baseline), Calibrate (returns the sensor value), Measure |
| Nephelometry | Re-calibrate (erases calibration), Calibrate with samples, Measure (results in NTU, plus a serial index that rises with each measurement) |
Calibration normally uses a formazin turbidity standard of 400 NTU diluted with distilled water, for example to 2, 4, 40, 100 and 200 NTU. The device reports that more time is needed until five values have been read.
The page ends with a warning: COD and turbidity are only two of many aspects of water quality, and depending on the questions being asked they may not be enough.
Sources
- Appropedia, Open-source mobile water quality testing platform, describing Bas Wijnen, G. C. Anzalone and Joshua M. Pearce, "Open-source mobile water quality testing platform," Journal of Water, Sanitation and Hygiene for Development 4(3), 532–537 (2014), doi:10.2166/washdev.2014.137; licensed CC BY-SA 4.0. Changed here: the text is rewritten and shortened, and the file lists, news links and the related-project thumbnails are left out; rewritten in hubnx's own words.
Comments